MAINTENANCE & CARE
6.13 Accessories
Type of couch
Art no:
RUBIN
Art no:
131435
Manual switch H
448582
●
Foot switch H
448583
●
Gas pressure spring
head section
402099
●
Gas pressure spring
footrest
435699
●
Kyphosis gas strut
401799
Mains cable PVC
Hanning
448584
●
Protective cover
600x400
190999
●
Accessories for couch type
7.0 Spare and wear parts
Repairs must not impair the safety of the couch. Therefore, they may only be carried out by the manufacturer or
persons authorized for this purpose. Only original spare parts from the manufacturer may be used for repairs.
If you notice any defects or wear on your therapy couch, please inform us using the contact details provided (see last
page). This also applies to type plates and labels.
When contacting us, always state the couch type, the article number and the serial number. This data can be found on
the type plate on the frame of the foot section end of the couch (see chapter 10).
Our service staff will be happy to assist you in defining the fault and in selecting the correct spare and wear parts such
as: Arm, foot, head or body pads, swivel castors, foot plugs, arm section guides as well as operating parts for the
electric height adjustment.
8.0 Service life, maintenance and care
The therapy couch you have purchased is designed for an extremely long service life. The limiting factor here is the
drive, which is designed for a lifetime of max. 10 years or 100000 cycles. After this time, depending on the condition of
the table, the drive can be replaced by our service department, provided regular inspection and maintenance.
8.1 Maintenance
In order to ensure the safety of the patient and user and to protect the therapy couch from damage, expert use and
regular maintenance in accordance with DIN 31051 are required. We recommend that you have the table serviced
every 1-2 years by professionally trained Schupp employees or by authorised specialist dealers. Within the scope of
the maintenance, all prescribed safety checks are carried out in accordance with current safety regulations:
►
Testing of medical technology according to DIN EN 62353 (VDE 0751-1), annually and in case of
changes
► BetriebSichV (Ordinance on Industrial Safety and Health)
►
MPBetreibV (Medical Devices Operator Ordinance)
DGUV (German Social Accident Insurance); Regulation 3 (formerly BGV A3)
carried out.
► Electrical voltage - keep your distance!
► Before carrying out any maintenance work, disconnect the recliners with electric motor from the
power supply by removing the mains plug of the motor connection cable from the mains socket!
Warning
.