Assembly and settings
10
01.00 | APS-ME1 | Assembly and Operating Manual | en | 1415570
4.2 Calibrating the electronic processor
Electronic processor
1
0-point
Voltage output
4
Amplification
Current output
2
Amplification
Voltage output
5
Activate voltage output
3
0-point
Current output
6
Activate current output
■
Gripper is mounted and connected.
■
Sensor is mounted and connected.
■
Electronic processor is mounted and connected.
■
Electronic processor is at operating temperature.
Ø
Set the standard signal output (0-10 V or 4-20 mA) at the DIP
switches. Make sure that the unused standard signal output is
deactivated.
Ø
Open the gripper. Insert the gauge block corresponding to the
minimum gripper stroke. Close the gripper.
Ø
Set the standard signal value corresponding to the inserted
gauge block at the 0-point potentiometer.
Ø
Open the gripper. Insert the gauge block corresponding to the
maximum gripper stroke. Close the gripper.
Ø
Set the standard signal value corresponding to the inserted
gauge block on the gain potentiometer.