3 Replacing the lamp
Please ensure that the lamp and lamp-
holder have cooled down before
replacement. The corresponding
warning symbol is attached to remind
you.
5 Troubleshooting
The display has a fault status indicator
(18). Any possible breakdowns can be
recognised quickly and easily.
„Err 1“:
Lamp circuit interrupted.
„Err 2“:
Temperature monitor has
triggered.
„Err 3“:
Short-circuit in the lamp
circuit, electronic fuse has
triggered.
First of all switch off the light source.
Open the lamp compartment (10) by
pressing the button (11) and pull it out
as far as the stop.
Press down the two levers of the speci-
al socket and pull out the faulty lamp.
The two levers must be pressed down
again while inserting the new lamp.
Push the lamp compartment in until it
latches (audible locking sound).
Switch the light source on.
4 Maintenance
Your KL 1500 LCD is maintenance-free.
There is no provision necessary for
disinfecting the light source when
using it in the medical field.
To clean the outside of the instrument,
use a soft dry cloth or commercially
available plastic cleaning cloths.
▼
▼
Lamp out,
fan not running,
no LCD display
Lamp out,
fan running,
fault status indication
„Err 1“
Lamp out,
fan running,
fault status indication
„Err 2“
Lamp out,
fan running,
fault status indication
„Err 3“
Instrument not switched on.
Plug not in socket.
No mains electricity voltage.
Lamp compartment not closed.
Fuse faulty.
Transformer overheated.
Lamp defective
No lamp
Insufficient cooling
Transient current increase
in lamp circuit.
Lamp causing short-circuit.
Switch instrument on.
Plug the plug in.
Check mains voltage.
Close lamp compartment.
Replace fuse (15).
Ensure adequate cooling, check
that lamp type is correct, re-start
instrument after cooling down
for a prolonged time.
Replace lamp (see point 3
of these instructions).
With light source switched
off, install lamp.
Ensure ventilation apertures are
free, avoid excessive ambient
temperatures, the instrument will
switch back on again after a
short time.
Switch instrument off and, after
a few seconds, back on again.
Replace lamp (see point 3
of these instructions).
Fault
Possible causes
Remedial action
If you are unable to rectify the fault by the actions mentioned above, please contact
your specialist dealer or the nearest SCHOTT agency. More extensive repairs must
be carried out by the authorised customer service depot.