handylab multi 12
What to do if...
71
6.4
General errors
Display LoBat
Instrument does not
react
to keystroke
Display
to
to
to
to
Probe symbol flashes
Message
StoFull
StoFull
StoFull
StoFull
Cause
Remedy
– Batteries almost empty
– Exchange the batteries (see
section 5.1 M
AINTENANCE
)
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Hold the <CAL> key
depressed and switch the
instrument on
Cause
Remedy
– Time-out of the interface
– Check that the instrument is
connected
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– All memory locations are full
– Output data storage and clear
data storage