4.3.6
Salt content correction
A salt content correction is required in the oxygen concentration mea-
surement of samples with a salt content of more than 1 g/l. To do this,
you have to enter the salinity equivalent of the test sample (range 0.0 -
70.0) and to switch on the salinity correction.
Note
With the handylab multi 12 handheld meter, you can measure the sa-
linity. How to proceed is described in section 4.4.3 M
EASURING
THE
SA
-
LINITY
on page 38.
Entering the salt con-
tent
Note
How to switch on the salt content correction is described on page 29.
1
Determine the salinity of the test sample (any method, see also
section 4.4.3 M
EASURING
THE
SALINITY
on page 38).
2
Press the <CAL> key repeatedly until Sal appears on the dis-
play.
3
Enter the salt content with <
> <
>.
4
Switch to the measuring mode with <M>.
°
C