Operation
19
Temperature measure-
ment in pH measure-
ments
You can perform pH measurements with or without a temperature
probe as well as with the temperature probe of an oxygen sensor or a
conductivity measuring cell. The measuring instrument recognizes
which sensors are connected and switches automatically to the correct
mode for the temperature measurement.
The following cases are distinguishable.
* If you do not wish that, you can:
– either disconnect the 2nd sensor and use the manual temperature
input or
– use an electrode with a temperature probe.
If a temperature probe is connected, it is indicated on the display by TP.
Note
When calibrating without a temperature probe (no TP display indicator
displayed), enter the current temperature of the respective buffer solu-
tion manually using the <
> <
> keys while keeping the <RUN/EN-
TER> key depressed.
Temperature probe
Display
Mode
pH
Cond or
Oxi
yes
-
TP
Automatic with
pH temperature probe
yes
yes
TP
-
-
Manual
-
yes
TP
flashes
The temperature value of the
second probe (Cond or Oxi) in
the same sample is taken
over for the pH measurement*