Instructions for Use SS 20.60
Page 6/16
Table of run-in and run-out distances
Flow obstacle
upstream
of the measuring conduit
Minimum length of
the run-in distance
(L1)
Minimum length of
the run-out
distance (L2)
Light bend (< 90°)
10 x D
5 x D
Contraction (conduit
contracts toward the
measuring area)
15 x D
5 x D
Expansion (conduit
expands toward the
measuring area)
15 x D
5 x D
90° bend or T-junction
15 x D
5 x D
Two 90° bends in one
plane
20 x D
5 x D
Two 90° bends with 3-
dimensional change in
direction
35 x D
5 x D
Shut-off valve
45 x D
5 x D
Table 1: Run-in and run-out distances
This table lists the
minimum values
required in each case. If the
listed straight conduit lengths cannot be achieved, the
measurement accuracy may be impaired
.