28
3.7.4 Memory mode “F”
Data is recorded during a user-defined period. The tension meter records
1000 measured values/sec.
3.7.4.1 Saving data
Prerequisites:
- In the main menu, select „Memory Settings“ and then memory mode F.
To save the first measurement series:
- Press the
button to start recording the measured values.
- While recording is in process, “Rec F” flashes in the display.
- Once the set time has elapsed, no more data will be saved
(the display shows “Mem F”, the memory position, and the free memory space).
- If you press the
button before the set time has elapsed, no more data will
be saved. The data is saved automatically (the display shows “Mem F”, the memory
position and the free memory space)
To abort data recording, press the
button. This deletes the procedure.
No data will be saved.
To save more measurement series:
- Press the
button to start recording the measured values.
- While recording is in process, “Rec F” flashes in the display.
fig. 3.7.4.1
3.7.4.2 Displaying the saved measured values
- Press the „Recall“ and buttons simultaneously to display the saved data.
- The and buttons allow you to switch between the various measurement series.
The display will only show statistical values:
- Number of batch (File)
- Name of the material to be measured
- Date and time of the measurement series
- Last reading
- Average
- Max.
- Min.
- Standard deviation (Std dev)
- Number of measured values (Records)
The measuring values can be shown as graph by pressing