33
3.7.6.4 Deleting the saved measured values
If data is saved in the tension meter, the display shows “Mem” and indicates the free
memory space.
Deleting data:
- Press the „Recall“ and buttons
- Then press the
button and confirm with the
button.
This clears the memory.
3.7.6.3 Displaying the saved measured values on a PC
The optional SCHMIDT software
Tension-Inspect 3
allows you easily and accurately
to analyze the saved measured values and export them to an Excel table.
Sample memory printout
03.02.2022 11:37
MTS-Series-306 200 cN
Unit : cN
Material: PA 0.2 mm
Start: 11:32:30 03.02.22
Stop: 11:32:38 03.02.22
Last: 23.3
Avg: 23.4
Max: 24.4
Min: 21.9
Std dev: 0.6
HI-Alarm: 85.0
Lo-Alarm: 15.0
Records: 6
22.6
24.4
23.5
21.9
24.1
23.9
4
PC Communication (USB Interface)
4.1
Installing the USB Driver
For data transfer to the PC, you will need a USB driver. If no matching driver is in-
stalled on your PC, the Windows Found New Hardware Wizard appears when you
connect the unit to the computer for the first time. A driver is also supplied with the op-
tionally available Tension Inspect 3 software.
The requirements of the CE specification are only complied with if the ten-
sion meter is equipped and operated with connecting cables supplied by
HANS SCHMIDT & Co GmbH. Certification to the CE specification does not
extend to, and shall be invalid for any other combination. Under no circum-
stances shall HANS SCHMIDT & Co GmbH be held liable for any damage re-
sulting from the use of non-SCHMIDT sensors or cables.