13
3.5.1 Recalling the Stored Tension Values
(Cont.)
The DISPLAY blinks, showing
26
Peak
cN
the maximum peak value of the measuring
period, the PEAK indicator and the symbol.
- Press the RECALL key.
The DISPLAY blinks, showing
19
Peak
cN
the minimum peak value of the
measuring period, the PEAK indicator and the symbol.
- Press the RECALL key.
The tension meter switches back to measuring mode.
The DISPLAY shows
3
cN
Mem
MEM and the current reading.
The tension meter is ready for a new measuring period. The previously stored tension
values will then be overwritten with new values.
3.5.2 Clearing the ETB Memory
If values are stored in the ETB memory, the DISPLAY shows e.g.
3
cN
Mem
with the
MEM indicator.
To clear the memory
:
- Simultaneously press the MEM and RECALL keys.
The DISPLAY shows e.g.
3
cN
; all values stored in the memory have been
deleted.
3.5.3 Memory Function HOLD
When the tension meter memory is empty, you can retain the last reading on the
DISPLAY by using the memory function HOLD.
To retain the last reading
:
- Press the RECALL / HOLD key once for about 1 second.
The DISPLAY shows
2 3 8
: .
cN
the last reading and the “:” colon symbol.
To switch back to measuring mode
:
- Press the RECALL / HOLD key once for about 1 second.
The tension meter switches back to measuring mode.
3.6 Error Messages
Error message 1
:
- The DISPLAY shows
EEE
.
The upper limit of the tension range was exceeded by more than 10%.
Reduce the line tension.
OR
AUTO ZERO is no longer possible.
Recalibrate the instrument following the directions in Chapter 3.9.