17
Material Setup
Submenu
Settings Menu
Description
No. of the cha-
racteristic mate-
rial curve 1 - 11
—
[character],
[numbers],
[special character]
to enter a name for the selected mate-
rial characteristic. The name can also
been adjusted by using the provided
software
Damping
—
[1] - [9];
[5]
Chapter 3.5.4
Alarms
—
[ON]
, [OFF]
To activate or deactivate the alarm
function for the material characteristic.
High limit
—
[0000] - [9999]
If the set limit value is exceeded, the
display reads
MAX-ALARM.
Low limit
—
[0000] - [9999]
If the value falls below the set limit
value, the display reads
MIN-ALARM.
Calibration
Chapter 3.6.1
Start
To perform a calibration, follow the in-
structions shown on the display.
Cal. Points
between [10 %]
and [100 %] of the
tension range
Set three calibration points for which a
calibration should be performed.
Weights
[cN/daN]
[g/kg]
Set the unit of the calibration weights
used.
3.4.1 Material menu
In the material menu you can make the settings for the selected material characteristics and
perform the calibration. To perform the calibration, the weights for the selected calibration
points must be available.
Factory settings are illustrated bold
3.4.2 Password
By setting a password, you can block the access to some menu parts for unauthorized
users:
0000
All menus are accessible.
0001 - 0999 The Material, Cal. Adjustment, Settings, and Memory Settings menus are
locked
1000 - 9999 The entire menu is locked.
To apply a changed password, exit the main menu.
Entering 0000 as the password disables the password feature.
If you forgot the password, please contact HANS SCHMIDT & Co GmbH to request
the master password. As an alternative, you can perform a factory reset. Please
note, that in this case, you will lose any customer-defined calibrations.