18.
Take a calibration weight. This must be at least 75 kilograms, place it on the animal cage
weigher
19.
If the display shows stable (top left), press the
button
20.
If display shows Pass in display then calibration is successful
21.
The computer restarts automatically
22.
Turn the computer off at the on/off button, and remove the weights.
23.
Turn the computer on again and weigh the weights. again to see if the correct weight is
displayed
12. Parameters
12.1 Parameters for MS EasyScale Pro Select 4309763
To set parameters, turn on the scale.
Press
key during the self-checking.
Display will show “pn”.
Press
,
and
to enter, display will show “po chk”.
Menu
Sub Menu
Description
P 0 chk
Set H
Set high limit for check weighing
Set lo
Set low limit for check weighing
beep
No
No beep for check weighing
Ok
Beep, when check weighing within limits
ng
Beep, when check weighing outside limits
P1 ref
A 2n 0
Automatic zero tracking
Options: 0 , .5d ,1 d , 2 d , 4 d
0auto
Zero setting range, after resetting the scale to zero.
Options: 0 , 2 , 5 ,10 , 20 , 50 , 100
0range
Manual zero setting range.
Options: 0 , 2 , 5 ,10 , 20 , 50 , 100
0tare
Auto zero tracking for NET mode.
(After pressing tare, NET weight will be zero)
Option: on / off
Speed
Set the ADC speed
Option: 7.5 / 15 / 3 0 / 60
Zero
Set new zero point
P 2 com
Mode
This option is used to set accumulation and RS-232
communication
Options:
Cont
: Continue data sending
St 1
: Send data once, when stable
St c
: Send data continuously, when stable
Pr 1
: Send data once, when print Key is pressed
(in printer mode)