4
Defibrillation
4.4
Automatic defibrillation
Page 36
A
rt.
no.
: 0-4
8-02
40 Rev. c
FRED PA-1
Step 3
Automatic shock delivery
As soon as the energy charging is completed, the device automatically delivers the
shock, without user intervention. An acoustic countdown starts and the orange button
blinks until the shock is delivered.
Shock hazard!
Do not, under any circumstances, touch the patient during shock delivery.
Make sure that the patient does not touch any conducting objects.
After the shock delivery, proceed with
Performing cardiopulmonary resuscita-
.
Step 4
Performing cardiopulmonary resuscitation
If the
FreeCPR
option is activated, the device instructs the rescuer to adjust the chest
compression frequency.
FreeCPR
measures the compression rate based on the impedance measurement by
the defibrillation electrodes.
5. Perform a CPR cycle. According to the configuration of the device, a CPR cycle
consists of:
– performing chest compressions for the set period of time, or
– alternately performing 30 chest compressions and 2 breathes for the set period
of time.
After the CPR cycle, the device continues automatically with
.
Finishing the
therapy