Electrode placement
4
User guide
Resting ECG with 10-lead patient cable
4.3
CARDIOVIT AT-1 G2
4.3 Resting ECG with 10-lead patient cable
Fig. IEC labelling
4.3.1
Electrode placement for standard leads
The electrode resistance can be checked in the electrode test screen (
see page 25
).
C1 red
R red
N black
C2 yellow
C3 green
C4 brown
C6 purple
C5 black
F green
L yellow
IEC label
AHA label
Connecting the ECG patient cable
C1 red
V1 red
Fourth intercostal space at the right sternal border
C2 yellow
V2 yellow
Fourth intercostal space at the left sternal border
C3 green
V3 green
Midway between sites C2 and C4
C4 brown
V4 blue
Fifth intercostal space on the mid-clavicular line
C5 black
V5 orange
Anterior axillary line on the same horizontal level as C4
C6 purple
V6 purple
Mid-axillary line on the same horizontal level as C4
L yellow
LA black
Left arm (resting ECG)
R red
RA white
Right arm (resting ECG)
F green
LL red
Left foot (resting ECG)
N black
RL green
Right foot (resting ECG)