Page 33
Maintenance
5
Operating Instruction
Changing the fuse and mains voltage
5.3
A
rt.
no.
: 2.
510
474 rev.
: d
ARGUS LCM/PLUS
5.3 Changing the fuse and mains voltage
Fuse types
V
The mains voltage may only be changed by qualified personnel.
V
Before changing the fuse and mains voltage, disconnect the device from the
mains and remove the mains plug. See section
V
Ruptured fuses must only be replaced with the fuse types indicated in the below
table.
Voltage range
Number
Fuse types
220 - 240 VAC
2
250 V 200 mA (T)
100 - 115 VAC
2
115 V 315 mA (T)
Changing the fuse
1. Disconnect the device from the mains and remove the mains plug. See chapter
2. Loosen the fuse using a screwdriver and remove it.
3. Replace both fuses (see "Fuse types" table).
4. Reinsert the fuse inset.
Changing the mains voltage
1. Disconnect the device from the mains and remove the mains plug. See section
2. Loosen the fuse using a screwdriver and remove it.
3. Remove the grey inset, turn it by 180° and reinsert it.
4. Check the voltage indication in the window.
5. Replace both fuses (see "Fuse types" table).
6. Reinsert the fuse inset.
Fig. 5.1
Fuse inset