Maintenance interval for the battery
Page 32
A
rt.
no.
: 2.
510
474 rev.
: d
ARGUS LCM/PLUS
5.2 Maintenance interval for the battery
• During normal operation no maintenance necessary.
• If not used every 3 months.
Replace the battery approx. every 4 years (depending upon application) if the actual
running time falls substantially under 30minutes.
5.2.1
Charging the battery
5.2.2
Battery disposal
Important
The battery is maintenance free during its normal life.
The battery should remain charged during storage. If the storage period exceeds
three months, recharge the battery.
Important
A totally discharged battery requires approx. 5 hours to be 90% recharged.
It is possible to use the unit when the battery is being charged. However, when this
is the case, the charging time of the battery will be substantially extended!
1. Connect the device to the mains but do not switch it on.
2. The LED for mains supply (1) is lit.
3. Charge the battery for at least 5 hours.
V
Danger of explosion! Battery may not be burned or disposed of domestic refuse.
V
Danger of acid burns! Do not open the battery.
The battery is to be disposed of in municipally approved areas or sent back to
SCHILLER AG.