Page 19
Operation
3
Operating Instruction
Initial settings
3.4
A
rt.
no.
: 2.
510
474 rev.
: d
ARGUS LCM/PLUS
3.4.4
Alarm limit setting
The following table gives the default alarm limits settings for adults. A changed val-
ue can be stored as default in the menu
About+/Software
.
The factory defaults are listed in the sections
and
.
All alarm limits are reset to the default or user specific system settings after switch-
ing the unit off/on, if they have not been stored as default. (see menu
About+/Soft-
ware
)
1. Press
alarm limit
button.
2. Press
up/down
or
left/right
buttons to select an alarm parameter and confirm
with
enter
. The entry field appears blue.
3. Press
up/down
to change the value and confirm with the
enter
button.
4. Press the
up/down
or
left/right
buttons to select other parameters or press the
alarm limit
button to exit the menu.
Alarm
Low
High
Unit
a
Prio.
a. "High" priority = audible signal with 2 x 5 impulses.
"Low" priority = audible signal with 1 x 2 impulses and 20 s pauses between the
impulse sequences.
b
b. Print "On" = Default printout containing a warning message when the min./max.
value is exceeded.
HR
50
140
/min
High
Off
ASYS
-
2
s
High
Off
c
RRi
c. If the RRi alarm option is
Off
in the
ECG
settings panel, the "Low" and "High" val-
ues for RRi and APNi will be "Off".
8
35
min
Low
Off
c
APNi
-
25
s
High
Off
SpO
2
90
101
%
Low
Off
PP
50
140
min
High
Off
SYS
100
140
mmHg
Low
Off
DIA
40
95
mmHg
Low
Off
MAD
70
140
mmHg
Low
Off
Ps
95
180
mmHg
Low
Off
Pm
50
100
mmHg
Low
Off
Pd
40
100
mmHg
Low
Off
eCO
2
35
45
mmHg
Low
Off
CO
2
i
0
2
mmHg
Low
Off
RRc
8
35
/min
Low
Off
APNc
-
25
s
High
Off