3.7
Procedure in case of an alarm
Page 24
A
rt.
no.
: 2.
510
474 rev.
: d
ARGUS LCM/PLUS
3.7.4
Preventive alarm suppression
The preventive alarm suppression is used to deactivate in advance all alarms that
may be caused by disconnecting patient cables, lose electrodes and relocation of
the patient.
3.7.5
Overview of physiological alarms
«
Press the
alarm
button and confirm with the
enter
button before an alarm is dis-
played.
Message
Alarm suppressed 3:00
is displayed. The time can be programmed in the
menu
About+/Alarm/Alarm suppr. time
.
The alarms are reactivated after the defined period of time has elapsed.
Removing the patient cables
When a cable is removed, the message
Cable off or no Sensor
is displayed.
«
Press the
alarm suppressing
button.
The alarm is deleted and the measurement and wave field is no longer displayed.
Alarm abbreviation
Description
Asys limit
Asystole time limit exceeded
SpO
2
Low/High
Oxygen saturation of the blood
PP Low/High
Peripheral pulse of SpO
2
RRi Low/High
Respiration rate impedance (from ECG electrode)
Apnea limit
Apnea time limit exceeded
CO
2i
low/high
Inspiratory CO
2
RRc low/high
Capnographic respiration rate
eCO
2
low/high
End-tidal expiratory CO
2
SYS low/high
Systolic pressure
DIA low/high
Diastolic pressure
MAD low/high
Mean atrial pressure
HR low/high
Heart reat
Ps low/high
Invasive systolic blood pressure
Pm low/high
Invasive mean blood pressure
Pd low/high
Invasive diastolic blood pressure
Temp low/high
When the temperature is outside the measuring range,
this is indicated by “<<“(15 below °C) or “>>” (above 45
°C). This limit is fixed.