22
Trouble shooting
If the unit malfunctions then please check the
following:
• Is the unit connected to a power supply (plug-in
mains supply unit or rechargeable battery pack)?
Check that the plug is correctly inserted.
• Has the ON/OFF button been pressed?
Press the ON/OFF button.
• Is the rechargeable battery pack charged?
If not, recharge the battery pack as described on
pages 20/21.
• Does the medicine chamber contain the correct
volume of medicine?
If not, fill the chamber with inhalation solution
as described in the Instructions for Use to ensure
that the nebuliser performance is optimal.
• Has the medicine chamber been treated with
grease or a detergent in error?
If so, rinse it out thoroughly with water that is
warm/hot (approx. 40°C).
• Has an air bubble developed in the ultrasonic
generator?
If so, remove it carefully using a blunt instrument.
• Has the overheating protection been triggered?
Check the quantity of medicine.
Allow the unit to cool for approx. 5 minutes.
• Has the unit turned itself off by reaching the resid-
ual volume or responds to temperature protec-
tion, to start at once the plug has to be removed
from the socket and has to be put in at once.