21
Setting the CO
2
value and pump pressure
During commissioning and after regular maintenance work, the CO
2
value must be set. Set the CO
2
value according to
the information on the burner at a boiler temperature of at least 60 °C.
For two-stage burners, both stages must be set.
To set the poop pressure, insert a pressure gauge into the measuring connection (P). Only use a sealing ring to seal
the pressure measuring connection. Do not use Teflon tape.
With the two-stage oil pumps, the pressure of the first and second burner stage is set separately on the oil pump via
the respective adjusting screws.
CO
2
–measuring connection
step 2
step 1
Pressure measuring
connection for pres-
sure gauge
pressure adjust-
ment step 2
pressure adjust-
ment step 1
Change the filter cartridge
of the oil filter (Item No. 040104) if the vacuum is
too high and
less than -0.3 bar.
Change oil filter
ATTENTION:
Please dispose of the oil filter or filter cartridge in an environmentally friendly manner.
A well adjusted burner has soot-free combustion. This means that the boiler requires little cleaning.
A thin, light grey layer
may be deposited in the combustion chamber. This is a sign of good combustion.
Do not remove this layer
as it acts like a preservative for the combustion chamber.
After you have dismantled the burner from the boiler, please follow these steps for cleaning the boiler.
Clean boiler
ATTENTION:
Korrosionsgefahr
Cleaning with liquids such as thinner or petrol and using brushes with metal bristles will cause corrosion.
Only use brushes or paintbrushes with plastic bristles for cleaning.
Do not use brushes or paintbrushes with metal bristles.
Sweep out and vacuum loose dust.
Item No.
bottle brush and corner
brush with soft bristles
090327
Make sure the room were the heater is located is well ventilated!
Take off the two nuts that hold the boiler door in place. Take off
the boiler door with the burner still attached. Follow the next
steps for cleaning the boiler.
1.
Remove the insulation plate on the front side of the boiler.
2.
Clean the tubes with a brush with synthetic bristles. Do not
use a metal brushes!
3.
Sweep the combustion chamber with a cleaning brush.
4.
Clean the front side of the boiler with a cleaning brush.
5.
Remove carefully any loose dust with a vacuum cleaner.
6.
Put the insulation plate back in its place.
7.
Reinstall the boiler door to the heater.
Summary of Contents for 077480
Page 31: ......