Advanced SkLineScan Software Functions
23
Instruction Manual
SK1024U3HU
shar
ed_CameraContr
ol(5)_Advanced-Sync-Ctrl_ML.indd
Instruction Manual SK1024U3HU © 2019-07 E
• Trigger on falling edge
of SyncA
• SyncB low active
• direction detection = on
• hysteresis = 0
Sync Control Register:
'1000 0xxx'b
B
1)
• Trigger on falling edge
of SyncA
• SyncB without function
• direction detection = off
• hysteresis = 0
Sync Control Register:
'0xxx xxxx'b
A
1)
• Trigger on falling edge
of SyncA
• SyncB low/high active
• direction detection = off
• hysteresis = 0
Sync Control Register:
'1100 0xxx'b
C
1)
• Trigger on 4 edges
of SyncA and SyncB
• direction detection = off
• hysteresis = 0
Sync Control Register:
'1110 0xxx'b
D
1)
• Trigger on 4 edges
of SyncA and SyncB
• direction detection = off
• hysteresis = 4
Sync Control Register:
'1110 1xxx'b
E
2)
1)
1)
1)
drive holdup oscillation
Example Timing Diagrams
Annotations:
SyncA
= LINE SYNC A (external line synchronization input, I/O connector)
SyncB
= LINE SYNC B (external line synchronization input, I/O connector)
Count
= internal counter
Trigger
= Generated trigger pulses from the Trigger Control stage. The signal goes to the Trigger
Divider stage inside the camera. For setting the divider, use the V
yyyyy
<CR> command or the
4.1 Camera Control by Commands (p. 20)
1) direction changed
2) glitch