PRE INSTALLATION GUIDE
WARNING: To ensure operator safety, radiation emission levels must be checked before putting
the SCANMAX into operation (See page 20)
Prior to first time operation, it is essential for the safety of the operator and for the long life of the
equipment that the following instructions are strictly observed.
Ensure that the system voltage is the same as the mains supply voltage available. If in doubt regarding the
mains voltage at hand, perform a measurement. Units will be set at 230V unless otherwise instructed and
should therefore only be operated from a 230V mains supply. Connecting to a different mains voltage will
result in improper operation or even destruction of the unit.
Ensure that a good mains earth is provided
. To minimise shock hazard the SCANMAX must be
connected to an electrical ground or earth. The unit is equipped with a three conductor AC mains lead.
The corresponding socket at the installation must be fitted with a reliable protective earth contact.
SCANNA or the supplier cannot be held responsible for incorrect connection.
Do not operate the equipment in the presence of flammable gases or fumes.
protective devices (fuses etc.).