DSA3017/3018 Firmware Replacement
Replacement of the Firmware chip should only be required if the DSA firmware is upgraded. It is not
expected that this chip might fail in normal use.
When software upgrades are available, customers will be notified by mail or FAX. Upgraded software will
only be made available to customers when requested. It is very important that the replacement procedure
be followed very carefully, otherwise damage to the chip or module may result. It is also very important to
observe ESD safety procedures while replacing this chip.
1
.
Remove the Splash Plate and the top two screws.
2. Place the module on its side as shown and
remove the two upper screws. Loosen but do not
remove the two lower screws.
3. Gently separate the housing, as shown.
4. Carefully remove the WSI chip with a PLCC
extraction tool (Scanco TL-108 or Techni-Tool
#560PR291). The use of any other tool will
damage the WSI chip.
5. Install the new chip. Be very careful to note the
proper orientation of the flattened corner. Push
firmly on the chip to seat it.
6. Reassemble the module in the reverse order. It
is very important that the wires be routed around
the shaded components.
32