SO-161-R Ophthalmic Microscope
User Manual
Page 13 of 18
Issue number 1.0
X:\Current User Manuals\SO-161-R User Manual v1.0.doc
STERILISATION AND CLEANING
1.
The removable covers may be sterilised by boiling, autoclaving, chemical sterilisation or
gas sterilisation.
2.
One set of covers can be sterilised while the other is in use.
3.
The anodised and plated metal components can be wiped with any of the normal
disinfectants.
4.
The plastic parts and the paintwork of the microscope assembly and the power supply may
be affected by organic solvents. Do not autoclave or wipe with organic solvents such as
ether, xylene or alcohol. To clean, use water-based solvents only.
5.
Focus covers are used to cover focus knobs only
6.
Zoom covers are used to cover the optional SO-1420 binocular assistant microscope
focusing knobs and SO-1510 PD adjuster knob
7.
Guide handle covers are used for the optional guide handle and side handles
CAUTION:
•
National authorities may require the use of specific sterilisation or disinfection methods.
Figure 9: Sterilisable covers