NOTE:
-
With approved instrument the accumulation functions with a minimum weight of 20d.
-
The maximum total value which may be viewed is 99999. Once reached these limits the message “- OF -” will
appear on the display.
-
The maximum total value which may be accumulated is 99999.
Once reached these limits, one should clear the total; by pressing
M+
the message “- OF -” will appear on the
display.
-
The total values will be cleared when it’s turned off.
6. SET-UP
6.1 NUMBER OF DIVISIONS AND ZERO TRACKING
Press together
PCS
and
7
keys at start-up, during self-checking:
-
The display shows the set number of divisions “
d
XXXX
”: press
TARE
to select 3000, 6000 or 7500, 15000, 30000
maximum divisions.
-
By pressing the
M+
key the display shows the set zero tracking value “
ZEr X.X
”, in other words, scale’s thermal
drift compensation parameter; the set value corresponds to the number of divisions which is cleared. Press
TARE
to select 0.0 (disabled), 0.5, 1.0, 1.5, 2.5 or 3.5.
-
Press
M+
to confirm.
6.2 CALIBRATION WITH NO ZERO ACQUISITION AND NO CAPACITY SELECTION
Press
MC
key till the weight window display “-CAL.-”, the unit weight window and the number window display 0 or
smaller number.
1.
F.S. calibration: Put enough weight on the platform. press
M+
when display is stable. The calibration finish.
2.
Calibrate by any weight: Press
TARE
, then weight window displays “-----“.
A
verage
P
iece
W
eight window and
number window doesn’t change. Input the weight by keyboard, then put the corresponding weight on the platform.
Press
M+
when display is stable. The calibration finishes.
Set the gravity of calibration and use zones (see section 6.4).
6.3 CALIBRATION WITH ZERO ACQUISITION AND CAPACITY SELECTION
1.
Turn off/on the scale.
2.
Press both the
PCS
and
8
keys during the turning on.
3.
By pressing the
TARE
key one selects the available scale capacity: 1.5, 3, 6, 15, 30, 60 kg. Confirm with
M+
.
4.
Unload the scale and press
M+
.
5.
Load a weight equal to the half capacity and press
M+
.
6.
Load a weight equal to the capacity and press
M+
.
7.
Set the gravity of calibration and use zones (see section 6.4).
6.4 GRAVITY OF CALIBRATION AND USE ZONES
a) Keep pressed the
TARE
key for about 4 seconds: the "Average Piece Weight" display will show "CAL_G", and the
"weight" display will show the gravity value of the calibration zone
b) Modify the value trough the numeric keyboard and confirm by pressing
M+.
c) The "unit weight" display will show "USE_G", and the "weight" display will show the gravity value of the use zone
d) Modify the value trough the numeric keyboard and confirm by pressing
M+.
e) Once the value is confirmed the scale return in weighing mode.
6.5 WEIGHING FILTER
4.
Press both the
PCS
and
5
keys during normal weighing.
5.
By pressing the
TARE
key one selects the available filter: SPd 0, SPd 1, SPd 2, SPd 3, SPd 4. The higher the
filter value, and greater is its intervention.
6.
Confirm with
M+
.