81
SBS-6500 User Manual
Storage Battery Systems, LLC.
14.2.
Fuse Replacement
[Warning]
To avoid electrical shock or bodily injury, remove any input signal from the test probes
before replacing the Fuse. Fuse should be replaced only with a proper rated fuse.
[Replacement Procedure]
①
Power OFF the SBS-6500.
②
Disconnect test probes from battery terminals.
③
Remove holster case.
④
Open tilt stand and remove the two battery compartment cover screws.
⑤
Unscrew the screw at lower area of bottom case.
⑥
Open the bottom of the case.
⑦
Remove Used Fuse and replace with new fuse.
⑧
Reassemble in the reverse order of above steps.
14.3.
Battery Replacement
[Warning]
To avoid electric shock or bodily injury, remove any input signal on test p robes before
replacing the batteries. Batteries should be replaced only with the same type of Battery
Packs supplied with the SBS-6500.
[Replacement Procedure]
①
Power Off the SBS-6500
②
Disconnect any test probes, test Leads, temperature probes, AC power adapter, USB
cable from the unit.
③
Remove holster case.
④
Open tilt stand and remove the screws from the two battery covers.