3) Selecting an Existing Entry from the Reference DB (Recall)
Recall an entry from the DB in the tester and use it to set limit values.
Desired DB entry can be selected by using the up/down
keys or encoder switch.
①
At limit set-up mode, press
to open the Reference DB.
②
Select desired DB by using up/down
key or encoder switch.
③
After DB entry is selected, press the
key to activate.
④
When the set-up or selection is complete, press the
key or encoder
switch to return to the measurement mode.
4.3.
Judgment of Limit
Measurements will be compared to the three programmable reference value alarm
thresholds. Based on the actual measurements, the Pass, Warning, or Fail status will
be displayed on the LCD, signaled by audible alerts and stored in the memory along
with the data from the measurements.
[How Results are Determined]
If the measured value of resistance is lower than warning level limit, and if voltage
measured simultaneously is higher than minimum limit then the condition is a
Pass
.
If the voltage measured is lower than minimum limit then the condition is a
Warning
with an audible alert.
If the measured value of resistance is higher than warning limit and lower than failure
limit or if voltage measured simultaneously is higher or lower than minimum V limit,
both cases the condition is a
Warning
with an audible alert.
If the measured value of resistance is higher than failure limit and if the voltage
measured simultaneously is higher or lower than minimum V limit, both cases the
condition is a
Failure
with an audible alert.