32 SBS-6000
below interface.
TEST SETTING
2011-05-10 09:01:23
1. String Mode:
7. Sample Test:
2. Cell Mode:
3. Conn Mode:
6. Set Test Type
:
4. Save Mode:
5. Set Warning:
Zero Adjust
Range
:
4mΩ
Function OFF
8. Zero Setting:
0.000
(
mΩ
)
Esc
Func On
Range
Func Off
Connect zero board
When tester is in impedance zero setting interfaces, it will alert “Please
connect to zero setting board” you have to connect testing clip or
probes to setting board, tester will be in auto zero setting. At the same
time impedance zero setting interface alert “Zero Setting” about 6
seconds to complete current gear zero setting, you can press F4key to
switch gear for each gear zero setting as below interface.
TEST SETTING
2011-05-10 09:01:23
1. String Mode:
7. Sample Test:
2. Cell Mode:
3. Conn Mode:
6. Set Test Type
:
4. Save Mode:
5. Set Warning:
Zero Adjust
Range
:
4mΩ
Function OFF
8. Zero Setting:
0.007
(
mΩ
)
Esc
Func On
Range
Func Off
Adjusting Zero