30 SBS-6000
2011-05-10 09:01:23
Sample Test
Esc
0 . 421 (mΩ)
R
Range
:
4 mΩ
Sample
:
30
REF Number
:
001
Sample NO.30
Save Ave
Save Min
Min
:
0.412(mΩ)
Average
:
0.432(mΩ)
5.3.8
Impedance Zero Adjustment Function
Functions description
This impedance zero functions is used to calibrate the tester by
adjusting impedance 0 range value, in order to make the test results
are much more reliable. When difference of environment temperature
and normal temperature 25
℃
is over 10
℃
or above this value, we
would suggest to use zero adjustment function. Correct and properly
use zero adjustment function that can improve measurement accuracy.
Note: In below description that mentioned zero adjustment board is in
good conductive metal plate, when zero adjusting, you have to try two
test clip or probe as close as possible.
How to use
Check below clip testing cable and setting board connection diagram.