39
Part No. 001052-EN, 03-m12-2019
Savaria PL Portable Ceiling Lift Owner’s Manual
Electromagnetic Immunity (continued)
Guidance and Manufacturer’s Declaration -
Electromagnetic Immunity - For All Equipment and Systems
The Monarch Ceiling Lift is intended for use in the electromagnetic environment specified below.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz,
1Khz, 80% AM
Mod.
10 V/m
80 MHz to 2.5
GHz, 1Khz, 80%
AM Mod.
Portable and mobile RF
communications equipment
should be used no closer to
any part of the lift, including
cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in metres.
Field strengths from fixed RF
transmitters, as determined
by en electromagnetic site
survey,
(a)
should be less than
the compliance level in each
frequency range.
(b)
Interference may occur in the
vicinity of equipment
marked with the following
symbol:
80 MHz to
800 MHz
800 MHz to
2.5 GHz
Summary of Contents for PL
Page 41: ......