36
Savaria PL Portable Ceiling Lift Owner’s Manual
Part No. 001052-EN, 03-m12-2019
7. TROUBLESHOOTING
Do NOT open the lift casing. Only an Authorized Savaria
Dealer is qualified to open it. Modifications made to the
lift by someone other than a qualified technician may
cause serious injury.
Problem
Action
The unit starts and stops
repeatedly.
If the load is greater that the Safe Working Load of
the lift, it will not function due to an overload on the
motor.
The lift emits a beeping
sound during use. The unit
may stop lifting the person
but can still lower the person.
The battery is low and needs to be charged.
Charging indicator on the
charger (yellow) does not
light up when the charger is
connected to the lift.
Ensure the charger is plugged into a standard
electrical outlet and that the outlet has power. The
green LED on the charger will be turned on.
Note that the battery may have sufficient power
and does not need recharging. The charger
automatically detects if charging is required or not.
The lift does not move when
you press a button on the lift
or on the hand control.
Make sure the On/Off power switch on the lift is
turned On.
Check that the battery is charged.
Ensure the hand control is properly plugged into
the lift socket.
Ensure the charger is not plugged into the lift. The
lift does not allow charging and lifting at the same
time.
If the problem persists and the lift still does operate,
contact your Authorized Savaria Dealer for service.
WARNING
Summary of Contents for PL
Page 41: ......