Instruction Manual 

HN-D 

 

HN-D-BA-e-1213 

 

2

 

 ( e.g. steel, aluminium etc.) the L-value represents a direct 
hardness  measurement  and  it  is  used  as  such. 
Comparison  curves  with  standard  static  hardness  values 
have  been  established  (Brinell,  Vickers,  Rockwell  C,  B, 
Shore D) for the most prevalent materials, enabling the L-
value  to  be  converted  into  the  relevant  values  for  these 
procedures.  With  HN-D,  such  hardness  values  can  be 
directly displayed in the hardness scales HLD, HRC, HRB, 
HB, HV, HSD. 
 

1.3 Features 
 

-  HN-D  ist  he  most  advanced  integrated  tester  (which 
means that the impact device D is integrated): no cables 
- Integrated impact direction sensor 
-  High  accuracy  (±  4  HL)  in  any  impact  direction  (360°)  – 
automatically 
-  Integrated  display  of  results  for  all  common  hardness 
scales 
-  Large  high-contrast    LCD  for  an  optimum  viewing  in  all 
conditions 
-Easy calibration 
- Full USB communication with PC, internal storage of data 
with day and time 
- Rechargeable Li-ion battery, charges through USB port 
- Intelligent sleep mode 
- Wireless (Bluetooth) to mini printer possible 
 

1.4 Applications 
 

- Convenient for all materials 
- Best suited for on-site testing of heavy, big or already    
   installed parts 
- Handy for difficult to accessor confined test locations 
- Automatic sensing and compensation for impact direction 
- Excellent for material selection and acceptance  tests. 
 

1.5 Primary Industries 
 

- Metal production & processing 
- Automotive & transportation 
- Machinery & power plants 
- Petro-chemical, refinery 
- Aerospace & shipyard 
- Metal constructions testing services & laboratories 
- Metallproduktion und Entwicklung 

 

2. Technical Specifications 
 

- Indication range:    0 to 999 HLD 
- Accuracy:    ± 4HLD  
- Measuring direction: any direction 
- LCD display: 128 x 64 dot LCD with backlight 
- Data memory: 500 groups 
- Measurement results can be automatically converted into 
  HRC, HRB, HB, HV, HSD 
- Impact energy: 11N 
- Weight of the impact body: 5,5 g 
- Impact tip: Diameter:   3mm 

                    Material:   tungsten carbide 
                    Hardness:   ≥ 1600 HV 
- Power: Li-ion rechargeable battery 
- Charger: DC 5V/500mA or USB port 
- Maximum continuous working time: approx. 16 hours 
- Operating environment: - Temperature:  -10°C to +60°C 
                                         - Humidity: 5% bis 95% 
- Dimensions: 147 x 35 x 22 mm 
- Weight: 63 g 
 

3.  Instrument’s views 
 

 

 
4.  Checking Supplied Accessories 

 
You’ll  have  to  check  in  advance  whether  you  obtained  all 
accessories  in  the  delivery  of  the  HN-D.  Other 
accessories, not included in delivery, can be disponible at 
any time by SAUTER GmbH. All items should be only used 
with the original , authorized instruments. The use of these 
accessories  with  other  instruments  could  cause  problems 
and any repair costs would not be covered by warranty. 
 

Packing list: 
 

Robust carrying case 

HN-D Metal hardness tester 

USB Cable 

Charger for HN-D 

Small support ring 

Cleaning brush 

Instruction Manual (CD) 

 

Note: The test block is not included in the delivery! 

 

5.         Operation Instruction 
5.1       Keys and their Functions 

                   

1.

“Next“

, select material, scale etc. can be usedR 

Summary of Contents for HN-D

Page 1: ...The environmental temperature should be within the range of 30 C up to 80 C and relative humidity from RH 5 bis 95 1 Summary 1 1 Measuring Principle Scope When the test is carried out an impact body with a tungsten carbide test tip is being impelled by spring force against the surface to be tested from which it rebounds Impact and rebound velocities are measured by the following method a permanent...

Page 2: ...xcellent for material selection and acceptance tests 1 5 Primary Industries Metal production processing Automotive transportation Machinery power plants Petro chemical refinery Aerospace shipyard Metal constructions testing services laboratories Metallproduktion und Entwicklung 2 Technical Specifications Indication range 0 to 999 HLD Accuracy 4HLD Measuring direction any direction LCD display 128 ...

Page 3: ...to be pressed to select the hardness scale The hardness scale changes according to the following sequence Illustration highlighted Hardness Scale character HLD Leeb Hardness HB Brinell HRB Rockwell B HRC Rockwell C HV Vickers HSD Shore Hardness D Note 1 The value means out of range 2 The default hardness scale is HLD 5 3 3 Data in One Group In measurement mode the button has to be pressed four tim...

Page 4: ...s can be browsed and by pressing the button the error values can be deleted 3 If the button is pressed it can be started to input the value written on the test block Primary the hundred bit has to be set see illustration below 4 By pressing the button it can be increased from 0 to 9 to set the hundred bit 5 By pressing the button the ten bit will be highlighted By pressing the button it can be inc...

Page 5: ...ker angeschaltet so ist nur die Taste zu betätigen um die laufenden Messdaten auszudrucken Mini Printer SAUTER AHN 01 www sauter eu A complete test report is shown below 7 Hardness Testing 7 1 Checking of Settings The button has to be pressed to turn on the instrument First it has to be checked if it has to be charged Then it has to be inspected if every setting is correctly particularly material ...

Page 6: ...tenance After 1000 2000 times use the impact device and the impact body should be cleaned with a nylon brush The screw of the supporting ring has to be turned off before the catheter can be cleaned Then the impact body has to be taken out and it has to be rotated into the tube with the nylon brush in anti clockwise direction The brush has to be pulled out when touching the bottom This procedure ha...

Page 7: ...ght of the sample is more than or equal to 5 kg testing can be directly started If it is less than 5 kg the sample needs a special charge The supporting piece has to be coupled by a coupling medium and pressed onto the surface of the test piece By doing this better test results will be achieved The area of the testing points should be free of vibrations or shaking If the weight of the sample is no...

Reviews: