Instruction Manual
HMR
HMR-BA-e-1212
12
Table 3
Type of impact device
DC(D)/DL
D+15
C
G
E
Impact energy
Mass of impact body
11mJ
5.5g/7.2g
11mJ
7.8g
2.7mJ
3.0g
90mJ
20.0g
11mJ
5.5g
Test tip hardness:
Diam. Test tip:
Material of test tip:
1600HV
3mm
Tungsten
carbide
1600HV
3mm
Tungsten
carbide
1600HV
3mm
Tungsten carbide
1600HV
5mm
Tungsten carbide
5000HV
3mm
synthetic diamond
Impact device diameter:
Impact device length:
Impact device weight:
20mm
86(147)/ 75mm
50g
20mm
162mm
80g
20mm
141mm
75g
30mm
254mm
250g
20mm
155mm
80g
Max. hardness of sample
940HV
940HV
1000HV
650HB
1200HV
Mean roughness value of
sample surface Ra:
1.6μm
1.6μm
0.4
μm
6.3μm
1.6μm
Min.weight of sample:
Measure directly
Need support firmly
Need coupling tightly
>5kg
2
~
5kg
0.05
~
2kg
>5kg
2
~
5kg
0.05
~
2kg
>1.5kg
0.5
~
1.5kg
0.02
~
0.5kg
>15kg
5
~
15kg
0.5
~
5kg
>5kg
2
~
5kg
0.05
~
2kg
Min. Thickness of sample :
Coupling tightly
Min. layer thickness for
surface hardening
5mm
≥0.8mm
5mm
≥0.8mm
1mm
≥0.2mm
10mm
≥1.2mm
5mm
≥0.8mm
Size of tip indentation
Hardness
300HV
Indentation
diameter
Depth of indentation
0.54mm
24
μm
0.54mm
24
μm
0.38mm
12
μm
1.03mm
53
μm
0.54mm
24
μm
Hardness
600HV
Indentation
diameter
Depth of indentation
0.54mm
17
μm
0.54mm
17
μm
0.32mm
8
μm
0.90mm
41
μm
0.54mm
17
μm
Hardness
800HV
Indentation
diameter
Depth of indentation
0.35mm
10
μm
0.35mm
10
μm
0.35mm
7
μm
--
--
0.35mm
10
μm
Available types of impact devices
DC: test hole or
hollow
cylindrical
DL:
test
slender
narrow groove or
hole
D
+15:
test
groove
or
reentrant
surface
C: test small, light
thin
parts
and
surface
of
hardened layer
G:
test
large,
thick, heavy and
rough
surface
steel
E:
test super high
hardness material