20
FS-BA-en-2210.docx
11.2.3 Functions
11.2.3.1 Track
Measured values are displayed continuously and stored with the set sampling
frequency if required.
To save the readings:
1. Press the Start/Stop button
2. The measured values are now saved and the generated file name in which the
data are saved appears in the display under the "Recording in progress"
indicator.
3. Pressing the Start/Stop key briefly stops the storage of the measurement series,
"Recording paused" appears in the display.
4. Pressing briefly again continues the recording. Recording in progress" is
displayed
5. Press the Start/Stop button for more than 3 seconds to end the recording in the
displayed file. Recording finished" display
6. To save a new measurement series in a new file, proceed again as in point 1.
11.2.3.2 Peak
To detect a peak value, select the Peak setting. When the measured value is shown
in the display, start the detection by briefly pressing the START/STOP button.
Please note that a peak can only be detected above the threshold value of +/- 5% of
the nominal measured value. For example 100N measuring range: Peak detection
from +5N or -5N.
The message "Peak running" appears in the display. Carry out your measurement. If
a peak value is reached and then falls again, the measurement is ended. You can now
see the file name of the measurement in the display and transfer it to your
measurement protocol.