11.2.1.2.2 For short time measurements
Such as maximum load measurement to destruction, which lasts only a few seconds
choose a higher frequency.
11.2.2 Readability
By clicking on change you can increase the readability by the displayed factor. The
maximum resolution is n=10,000
The smallest adjustable readability for a sensor is calculated as follows:
𝑑
𝑚𝑖𝑛
: 𝑘𝑙𝑒𝑖𝑛𝑠𝑡𝑒 𝐴𝑏𝑙𝑒𝑠𝑏𝑎𝑟𝑘𝑒𝑖𝑡 𝑖𝑛 𝑁
𝑀𝐴𝑋: 𝐾𝑒𝑛𝑛𝑤𝑒𝑟𝑡 𝑖𝑛 𝑁
𝑛: 𝑚𝑎𝑥𝑖𝑚𝑎𝑙𝑒 𝐴𝑢𝑓𝑙ö𝑠𝑢𝑛𝑔
𝑑: 𝑒𝑖𝑛𝑔𝑒𝑠𝑡𝑒𝑙𝑙𝑡𝑒 𝐴𝑏𝑙𝑒𝑠𝑏𝑎𝑟𝑘𝑒𝑖𝑡 𝑖𝑛 𝑁
𝑆𝑇𝐸𝑃: 𝐴𝑏𝑙𝑒𝑠𝑏𝑎𝑟𝑘𝑒𝑖𝑡𝑠𝑓𝑎𝑘𝑡𝑜𝑟
𝑑
𝑚𝑖𝑛
=
𝑀𝐴𝑋
𝑛
𝑑 = 𝑑
𝑚𝑖𝑛
∗ 𝑆𝑇𝐸𝑃 =
𝑀𝐴𝑋
𝑛
∗ 𝑆𝑇𝐸𝑃
Example FS 2-100:
𝑀𝐴𝑋 = 100 𝑁
𝑑
𝑚𝑖𝑛
=
100𝑁
10.000
= 0,01𝑁
𝑆𝑇𝐸𝑃 = 2
𝑑 =
𝑀𝐴𝑋
𝑛
∗ 𝑆𝑇𝐸𝑃 =
100𝑁
10.000
∗ 2 = 0,02𝑁
Set the readability factor as high as possible and as low as necessary! Very low
readability often leads to unstable display values with the measuring technology used
and the high sampling frequencies. Readability factor 1 should only be set if the
measured data is subsequently processed with additional PC programs, for example
to filter, smooth, etc. the stored values.
If values are only read or stored manually, it is recommended to set the readability
factor ≥ 5.