‘Down sensor’ connector from the value of the magnetic field of the sensor connected to 
the ‘Up sensor’ connector. 

 If the position on a grid should be used, the position of the first measurement 

must be set, which means setting the items ‘Line’, ‘Pos’, and ‘Step’ in the POSITION item 
of the main menu. This is described in detail in chapter 5.7. The measurement is started 
by pressing [ENTER]. After about 2 seconds, the measured value of the magnetic field 
with the statistical error is displayed, as well as the position of the measurement point 
‘Line’ and ‘Pos’ and other information. 

     S 08:04:28                 

        46347.2

  

±05.6nT

 

        

Line:+1026 Pos:-0980 

               Amp:8 D:0.6s 
 

               Save? ENTER/ESC 

The value Amp means the signal intensity, and the value D expresses a time 

constant of the decay of the precession signal. The question at the bottom indicates the 
option to store the reading in the data memory.  

By pressing [ENTER], the reading is stored in the data memory, and the next 

position is displayed. By pressing [ESC], the measured data is erased, and the current 
position is displayed. 

6.5. 

Measuring In Walk Mode 

The WALK measure mode is selected in the MEASURE menu. In this mode, the 

instrument is prepared to measure int the single and or gradient 

of the Earth’s magnetic 

field based on sensor configuration. The sensor configuration can be selected in the 
setup menu item MODE see chapter 5.4. 

In the WALK measure mode, the measurements are continually repeated and 

saved to memory. The repetition interval is selectable int setup menu item WALK MODE. 
The intervals are 0.5s, 1s, and 1.5s see chapter  

6.6. 

Measuring In Auto Mode 

The AUTO measure mode is selected in the MEASURE menu. In this mode, the 

instrument is prepared to serve as a base station and measure diurnal variations. When 
the 

measurement is started, two items are displayed: ‘Start time’ sets the time of the first 

measurement of the base station.  

‘Interval’ sets the time interval between measurements. 

The results of the measurement in the AUTO mode are stored automatically in the 

data memory. 

The measurement in the AUTO mode is terminated by switching the instrument 

off. 

Summary of Contents for PMG-2

Page 1: ...2021 SatisGeo s r o 1 29 2021 PMG 2 User Manual...

Page 2: ...12 5 2 Tuning Of The Magnetometer 12 5 3 Setting The Backlight 14 5 4 Selection Of The Sensor Configuration 14 5 5 Walk Sample Rate Configuration 14 5 6 Setting The Time Interval To Switch Off 14 5 7...

Page 3: ...2 8 1 USB Interface 19 8 2 Charger 19 8 3 Maintenance And Repair Work 19 8 4 Storage And Transportation 19 8 5 Warranty 19 8 6 Accessories 20...

Page 4: ...ield WALK mode continually measures the magnetic field based on sensor configuration in three intervals 0 5s 1s 1 5s 2s 2 5s and 3s The possible sensor configuration is the same as in manual mode The...

Page 5: ...ght 3 2 kg including battery Sensor Dimensions dia 80 x 200 mm Weight 0 7 kg Operating temperature range 10 to 60 C Storage temperature range 20 to 70 C The maximum gradient 1000nT m means that if a c...

Page 6: ...l from the sensor and thus the size of the error of the measurement Figure 3 1 Post assembly for line measurement For measurements the sensor is mounted on a post which is assembled using up to five h...

Page 7: ...ue of the field measured in the location of the sensor connected to the Up sensor connector minus the value of the field measured in the location of the sensor connected to the Down sensor connector T...

Page 8: ...proportional to the measured magnetic field The entire measuring cycle is controlled by a microprocessor which computes the measured magnetic field and statistical error of the measurement from the va...

Page 9: ...parameters 1 MEASURE item sub menu serves to activates measurement The possible measurements are MANUAL WALK and AUTO See chapter 6 2 POSITION Item serves for the setting of a number of a profile line...

Page 10: ...ated in the setup menu item 5 S 08 04 28 DEVICE SETUP 1 BACKLIGHT 2 DATE TIME 3 OFF TIME 4 INFO 1 BACKLIGHT Item serves to switch the backlight on off or to set the time to elapse See chapter 5 3 2 DA...

Page 11: ...10...

Page 12: ...ivities of the operator on the keyboard or through the USB interface The interval can be set in the Setup menu item Time If the instrument does not display the above mentioned information even after r...

Page 13: ...d magnetic field To display the current tuning interval select the TUNE sub menu MANUAL SET item in the setup menu For the SINGLE mode the display shows for example S 08 04 28 SET TUNE Sensor UP 41600...

Page 14: ...00 43 800 nT 35 95 300 99 999 nT 18 43 500 45 800 nT Table 5 1 Tuning intervals A preliminary approximate setting of the tuning point can be performed on the basis of a magnetic field map see Figure 5...

Page 15: ...ttery the instrument is automatically switched off after elapsing the set time interval after the last activities of the operator on the keyboard Thus the deep discharge and the damage of the battery...

Page 16: ...and is changed by pressing 0 when the cursor is placed at the sign NEXT is used for switching among the items Line Pos and Step The modified data are stored in the memory by pressing ENTER Pressing ES...

Page 17: ...sensor is changed and the statistical error of the measurement increases To achieve the maximum signal the sensor must be oriented in the north south direction so the north mark on the sensor must po...

Page 18: ...displayed By pressing ESC the measured data is erased and the current position is displayed 6 5 Measuring In Walk Mode The WALK measure mode is selected in the MEASURE menu In this mode the instrument...

Page 19: ...mation To store the data the operator presses ENTER otherwise ESC canceling the measured value and the measurement on the point can be repeated The position of the next measurement is displayed in bot...

Page 20: ...netometer requires no special maintenance Only the built in lead acid battery needs care If the voltage drops under a certain level the battery might be damaged or the magnetometer might not operate c...

Page 21: ...cessing Unit Console 1 pcs Sensor 2 pcs Harness 1 set Foot Stick 1 pcs Middle Stick 4 pcs Anchor Ring with Ropes 1 set Tent Pin 3 pcs Interface Cable to PC 1 pcs Charger 1 pcs Instruction Manual 1 pcs...

Reviews: