background image

16 

 

6. MEASUREMENT 

The instrument and its accessories are transported in a case. It must be 

assembled before an operation. 

The measured value of the magnetic field depends very much on the presence of 

metal objects near the sensor. The operator should remove all the metal objects prior to 
measuring. 

6.1. 

Instrument Setup 

Screw together the sensor and the sticks of the post to the desired length 

– see 

Figure 3.1. The processing unit can be attached to the harness and carried on the chest. 

Connect the sensor to the processing unit by using the appropriate connectors. In 

the single sensor configuration and AUTO modes, the connecto

r ‘Up sensor’ is used. In 

the GRAD sensor mode, both sensors are used. The upper sensor should be connected 
to the ‘Up sensor’ connector otherwise the sign of the gradient value is changed. 

It is recommended to keep the sensor as far as possible from the processing unit 

to limit the influence of the magnetic moment of the instrument. 

For the same reason, it is recommended to keep the direction of the advance 

along with the profile during the measurement in one area. 

6.2. 

Sensor Orientation 

The measured value of the magnetic field does not depend on the sensor 

orientation. However, 

with the changing orientation of the sensor towards the Earth’s 

magnetic field, the signal from the sensor is changed, and the statistical error of the 
measurement increases. To achieve the maximum signal the sensor must be oriented in 
the north-south direction, so 

the ‘north’ mark on the sensor must point to the geographical 

north. 

6.3. 

Tuning 

Before measuring in a new area, it is necessary to set or check the preliminary 

tuning of the instr

ument according to the map of the Earth’s magnetic field, see Figure 

5.1. If the measurement is conducted with inaccurately tuned circuits, an error message 
is displayed. If the automatic tuning function is on, the tuning will occur automatically 
concerning the measured value. The automatic tuning is not performed in the case of an 
enormous difference in which the value of the magnetic field is measured with a 
significant statistical error. The tuning procedure is described in chapter 5.2. 

6.4. 

Manual Measuring 

MANUAL measuring is started by selectin MANUL in the measure menu. The 

selection is asserted by [ENTER]. The instrument is prepared to measure in the single 
and or gradient sensor configuration. The sensor configuration can be selected in the 
setup menu item MODE see chapter 5.4. In the single sensor configuration, the 
instrument measures the absolute value of the magnetic field vector. In the gradient 
sensor configuration, the device measures the value of the magnetic field in both 
locations of the sensors simultaneously. The displayed value of the gradient is 
determined by subtracting the value of the magnetic field of the sensor connected to the 

Summary of Contents for PMG-2

Page 1: ...2021 SatisGeo s r o 1 29 2021 PMG 2 User Manual...

Page 2: ...12 5 2 Tuning Of The Magnetometer 12 5 3 Setting The Backlight 14 5 4 Selection Of The Sensor Configuration 14 5 5 Walk Sample Rate Configuration 14 5 6 Setting The Time Interval To Switch Off 14 5 7...

Page 3: ...2 8 1 USB Interface 19 8 2 Charger 19 8 3 Maintenance And Repair Work 19 8 4 Storage And Transportation 19 8 5 Warranty 19 8 6 Accessories 20...

Page 4: ...ield WALK mode continually measures the magnetic field based on sensor configuration in three intervals 0 5s 1s 1 5s 2s 2 5s and 3s The possible sensor configuration is the same as in manual mode The...

Page 5: ...ght 3 2 kg including battery Sensor Dimensions dia 80 x 200 mm Weight 0 7 kg Operating temperature range 10 to 60 C Storage temperature range 20 to 70 C The maximum gradient 1000nT m means that if a c...

Page 6: ...l from the sensor and thus the size of the error of the measurement Figure 3 1 Post assembly for line measurement For measurements the sensor is mounted on a post which is assembled using up to five h...

Page 7: ...ue of the field measured in the location of the sensor connected to the Up sensor connector minus the value of the field measured in the location of the sensor connected to the Down sensor connector T...

Page 8: ...proportional to the measured magnetic field The entire measuring cycle is controlled by a microprocessor which computes the measured magnetic field and statistical error of the measurement from the va...

Page 9: ...parameters 1 MEASURE item sub menu serves to activates measurement The possible measurements are MANUAL WALK and AUTO See chapter 6 2 POSITION Item serves for the setting of a number of a profile line...

Page 10: ...ated in the setup menu item 5 S 08 04 28 DEVICE SETUP 1 BACKLIGHT 2 DATE TIME 3 OFF TIME 4 INFO 1 BACKLIGHT Item serves to switch the backlight on off or to set the time to elapse See chapter 5 3 2 DA...

Page 11: ...10...

Page 12: ...ivities of the operator on the keyboard or through the USB interface The interval can be set in the Setup menu item Time If the instrument does not display the above mentioned information even after r...

Page 13: ...d magnetic field To display the current tuning interval select the TUNE sub menu MANUAL SET item in the setup menu For the SINGLE mode the display shows for example S 08 04 28 SET TUNE Sensor UP 41600...

Page 14: ...00 43 800 nT 35 95 300 99 999 nT 18 43 500 45 800 nT Table 5 1 Tuning intervals A preliminary approximate setting of the tuning point can be performed on the basis of a magnetic field map see Figure 5...

Page 15: ...ttery the instrument is automatically switched off after elapsing the set time interval after the last activities of the operator on the keyboard Thus the deep discharge and the damage of the battery...

Page 16: ...and is changed by pressing 0 when the cursor is placed at the sign NEXT is used for switching among the items Line Pos and Step The modified data are stored in the memory by pressing ENTER Pressing ES...

Page 17: ...sensor is changed and the statistical error of the measurement increases To achieve the maximum signal the sensor must be oriented in the north south direction so the north mark on the sensor must po...

Page 18: ...displayed By pressing ESC the measured data is erased and the current position is displayed 6 5 Measuring In Walk Mode The WALK measure mode is selected in the MEASURE menu In this mode the instrument...

Page 19: ...mation To store the data the operator presses ENTER otherwise ESC canceling the measured value and the measurement on the point can be repeated The position of the next measurement is displayed in bot...

Page 20: ...netometer requires no special maintenance Only the built in lead acid battery needs care If the voltage drops under a certain level the battery might be damaged or the magnetometer might not operate c...

Page 21: ...cessing Unit Console 1 pcs Sensor 2 pcs Harness 1 set Foot Stick 1 pcs Middle Stick 4 pcs Anchor Ring with Ropes 1 set Tent Pin 3 pcs Interface Cable to PC 1 pcs Charger 1 pcs Instruction Manual 1 pcs...

Reviews: