SATEC PIPE & PRESSURE
24
To perform an
FM burst test
:
1.
Press the Page button until “
Prog
” is displayed.
2.
Press the Scroll button until “
tGt
” is displayed. This is the target pressure
value in psi. Use the Up/Down buttons to adjust the value as desired. The
display will blink to show that the new value is accepted.
3.
Press the Scroll button until “
dur
” is displayed. This is the test duration in
seconds. Use the Up/Down buttons to adjust the value as desired. The display
will blink to show that the new value is accepted.
4.
Press the Scroll button until segment 2 is displayed. Press the Scroll button
again to display the current parameter type. Use the Up/Down buttons to
adjust the type to “
dwEll
”.
5.
Press the Scroll button until “
dwEll
” is displayed in the upper readout. This is
the dwell time in seconds. Use the Up/Down buttons to adjust the value as
desired.
6.
Press the Scroll button until segment 3 is displayed. Press the Scroll button
again to display the current parameter type. The parameter type should be
“
End
”. If it is not “
End
”, use the Up/Down buttons to adjust the type to
“
End
”.
7.
Press the Scroll button to check the setpoint value for the “
End
” parameter. It
should be “
rSEt
”.
8.
Press Page and Scroll simultaneously to return to the
Home
display.
NOTE:
You can get back to the
Home
display any time by pressing the Page and
Scroll buttons simultaneously. Also, the
Home
display will return if no
button is pressed for 45 seconds (10 seconds if the controller is indicating
an alarm condition) or whenever the power is turned on.
Temporary Adjustment of Setpoint
When a program is not running, the setpoint can be temporarily adjusted by
pressing the Up or Down buttons as appropriate. This does not change the
programmed setpoint for a test.