![Satec 3100 Series User Manual Download Page 18](http://html.mh-extra.com/html/satec/3100-series/3100-series_user-manual_1200394018.webp)
Series 3100
PP3123-03-0101
Automatic Burst Test Systems
User’s Guide
SATEC PIPE & PRESSURE
18
High Pressure Relief Valve
A high pressure relief valve protects the system from overpressure by relieving
excess pressure. The relief valve discharges excess water from a tube on the side
of the control cabinet. If water drainage is visible from this tube, immediately
disconnect the air supply and slowly open the by-pass valve until pressure is
below the system maximum; watch the
SUPPLY PRESSURE
gauge. Call
INSTRON-SATEC SYSTEMS for service.
WARNING!
Immediate attention is needed if this valve starts relieving high pressure.
Shut off the air to the system and bleed off the high pressure by opening
the internal by-pass valve inside the cabinet. Check system thoroughly,
especially the pump’s air regulator for correct setting, before returning the
system to service. Call INSTRON-SATEC SYSTEMS for service.
Multiple Specimen Option
A multiple specimen option can be purchased that permits the preparation and
testing of multiple specimens. Specimens are tested one at a time through the use
of a
SPECIMEN
selector dial. The number of specimens that can be prepared
and tested varies from system to system and depends on customer requirements.