Printed on 26/04/12 13/15 U514986-e Revision 4
5 - Check adjustment by carrying out a new gauging test
6 - Put the cover back in place and tighten the screws.
7 - Re-seal
C - PRINCIPLE OF ADJUSTMENT
Reminder : 1 notch of the calibration screw (3) = 0.25‰ in whatever direction adjustment is made.
1 - Note the position of the calibrating screw.
2 - Turn the screw the appropriate number of notches in the direction required to obtain the required
adjustment :
difference (in ‰) / 0.25 = number of notches
In the above example, 3 litres must be added to the gauge. Therefore we must turn the calibration screw
(3) in the positive direction (+) of :
3 (‰) / 0.25 = 12 notches, in the anticlockwise direction.
D – ADJUSTMENT OF THE ELECTRONIC CALCULATOR RUBIS OR EQUALIS
See calibration instructions U513237 For the Rubis
See calibration instructions U516318
For the Equalis L
See calibration instructions U516703
For the Equalis MPC