4

Connect the scale to AC
power.  Install the power cable
and grounding cable with
adequate protection so that
they cannot be damaged.

Remove the load plate.
Use a 13-mm open-end wrench
to loosen the leveling foot (1).
Turn the leveling foot (2) until
the scale is level.  Tighten the
lock nut (1) with the wrench.
Replace the load plate

Ground the scale in the
hazardous area/location.
Connect an equipotential
bonding conductor to the
terminal (11).

After connecting the scale to AC power, wait approx. 30 minutes
for it to warm up.

Operating the Scale

Turn on the scale using the 

e

key (2).

After the scale has been turned
on, it will automatically run a
self-test.  At the end of this test,
0.0 g is displayed.

If a different readout is
displayed, zero or tare the
scale using the 

=

 key (4).

Summary of Contents for Sartorius PMA 35D-X

Page 1: ...98648 008 38 Sartorius PMA 35D X PMA 35D X00C PMA 35D X00CV1 Electronic Paint Mixing Scales Installation and Operating Instructions...

Page 2: ...Note to Users G Make sure to carefully read and follow sections marked with this symbol They contain important safety instructions and information G Use only the adapter that is included in delivery...

Page 3: ...09308 AC adapter must be installed outside the hazardous area location G Before you operate your scale in a hazardous area location the scale must be inspected either by a certified electrician or und...

Page 4: ...ut 1 with the wrench Replace the load plate Ground the scale in the hazardous area location Connect an equipotential bonding conductor to the terminal 11 After connecting the scale to AC power wait ap...

Page 5: ...before each calibration adjustment allow the scale to warm up for approx 30 min Hold down the key 4 for 2 sec When 5000 is displayed release the key Center the calibration weight on the load plate 1 C...

Page 6: ...00C Only Let s suppose that you poured in too much of one color component for a given formula e g with 4 components In addition let s assume that you previously poured in all of the other amounts exac...

Page 7: ...in too much The correct weight for the formula is 200 0 g 8 Press the y key 10 to start the recalculation mode A C correct flashes on the display 9 Press the y key 10 to correct the value to 200 0 g...

Page 8: ...able conditions 1 1 4 To obtain a detailed list of the menu codes please ask your nearest Sartorius office dealer or paint supplier Care and Maintenance Cleaning Use only naphtha or alcohol to clean t...

Page 9: ...ill of paint Please enclose a description of the equipment failures faults C Marking Note G The seal marking affixed to the equipment indicates that the equipment may be opened and serviced only by te...

Page 10: ...40 Power consumption VA average 8 Interface RS 232C Format 7 bit ASCII 1 start bit 1 or 2 stop bits Parity odd even mark or space Transmission rates 150 to 9600 baud Handshake Software or hardware Po...

Page 11: ...11...

Page 12: ...12...

Page 13: ...13...

Page 14: ...14...

Page 15: ...15...

Page 16: ...artorius AG The status of the information specifications and illustrations in this manual is indicated by the date given below Sartorius AG reserves the right to make changes to the technology feature...