9

Troubleshooting Guide

Problem

   Causes

   Solution

No segments appear

– No AC power available

– Check the AC power supply

on the weight display

The display shows “

L

– The load plate is not in place

– Position the load plate

The display shows “

H

– The load on the pan exceeds

– Unload the scale

   the scale’s capacity

The weight readout

– Unstable ambient conditions

– Set up the scale in another area

changes constantly

– Too much vibration or the

– adapt the scale to the particular

   scale is exposed to a draft

   weighing environment

The weight readout

– Scale not adjusted

–- Adjust scale

is obviously wrong

– The scale was not tared

   Tare/zero before weighing

   (zeroed) before weighing

Storage and Shipping Conditions

Storage temperature: – 40 to + 70°C/ – 40 to +158°F
Read and follow the instructions given in the section entitled “Safety Inspection.”If you need to return this
equipment, disconnect all cables before shipping to prevent damage.  Remove any splashes or spill of
paint.  Please enclose a description of the equipment failures/faults.

C

 Marking

Note:

G

 The seal marking affixed to the equipment indicates that the equipment may be opened and serviced

only by technicians authorized by Sartorius to ensure proper functioning and safe operation of this
equipment.  Otherwise, you will invalidate the warranty.

The equipment meets the requirements of the following Council Directives:

89/336/EEC “Electromagnetic Compatibility (EMC)”

Applicable European Standards:

Limitation of emissions:

 EN 50081-1

Residential, commercial and light industry

 EN 50081-2

Industrial environment

Defined immunity to interference:  EN 50082-1

Residential, commercial and light industry

 EN 50082-2

Industrial environment

Note:

The operator shall be responsible for any modifications to Sartorius equipment and for any connections of
cables or equipment not supplied by Sartorius and must check and, if necessary, correct these
modifications and connections.  On request, Sartorius will provide information on the minimum operating
specifications (in accordance with the Standards listed above for defined immunity to interference).

Summary of Contents for Sartorius PMA 35D-X

Page 1: ...98648 008 38 Sartorius PMA 35D X PMA 35D X00C PMA 35D X00CV1 Electronic Paint Mixing Scales Installation and Operating Instructions...

Page 2: ...Note to Users G Make sure to carefully read and follow sections marked with this symbol They contain important safety instructions and information G Use only the adapter that is included in delivery...

Page 3: ...09308 AC adapter must be installed outside the hazardous area location G Before you operate your scale in a hazardous area location the scale must be inspected either by a certified electrician or und...

Page 4: ...ut 1 with the wrench Replace the load plate Ground the scale in the hazardous area location Connect an equipotential bonding conductor to the terminal 11 After connecting the scale to AC power wait ap...

Page 5: ...before each calibration adjustment allow the scale to warm up for approx 30 min Hold down the key 4 for 2 sec When 5000 is displayed release the key Center the calibration weight on the load plate 1 C...

Page 6: ...00C Only Let s suppose that you poured in too much of one color component for a given formula e g with 4 components In addition let s assume that you previously poured in all of the other amounts exac...

Page 7: ...in too much The correct weight for the formula is 200 0 g 8 Press the y key 10 to start the recalculation mode A C correct flashes on the display 9 Press the y key 10 to correct the value to 200 0 g...

Page 8: ...able conditions 1 1 4 To obtain a detailed list of the menu codes please ask your nearest Sartorius office dealer or paint supplier Care and Maintenance Cleaning Use only naphtha or alcohol to clean t...

Page 9: ...ill of paint Please enclose a description of the equipment failures faults C Marking Note G The seal marking affixed to the equipment indicates that the equipment may be opened and serviced only by te...

Page 10: ...40 Power consumption VA average 8 Interface RS 232C Format 7 bit ASCII 1 start bit 1 or 2 stop bits Parity odd even mark or space Transmission rates 150 to 9600 baud Handshake Software or hardware Po...

Page 11: ...11...

Page 12: ...12...

Page 13: ...13...

Page 14: ...14...

Page 15: ...15...

Page 16: ...artorius AG The status of the information specifications and illustrations in this manual is indicated by the date given below Sartorius AG reserves the right to make changes to the technology feature...

Reviews: