Use in Combination with Other Programs
Practical Example: Checking Piece Count
Symbol displayed:
+/–
and “Counting” symbol:
P
Application: Checking the piece count of bulk parts of identical weight;
e.g. checking packages of screws where the packages must have
the same weight; checking the fill quantity; sorting.
Menu codes used in the example:
Function
Code
Output of application data: on 4 2 1
Output all parameters
7 1 2
Step/Key
Display
Data Output
Place empty container
on the scale
+
31.9 g
cs
Press the
=
/
z
key
0.0 g
cs
Press
g
/
W
to select
the reference sample quantity
rEF 20 pcs
Place 20 pieces
in the container
+
17.5 g
cs
Press
u
/
w
, activate the
counting application
+
20 pcs
The current piece count
and current reference weight
nRef
+
20 pcs
are output
wRef
+
0.8750 g
Fill-in required quantity
+
200 pcs
Press the
g
/
W
key +
200 pcs
Press the
z
/
T
key,
activate checkweighing
+
200 pcs
The tolerance limit symbol
flashes; the user guide display
shows
SEt = target weight
Press the
z
/
T
key to store
the weight value
+
200 pcs
The user guide display shows
S– – = d –
min
. weight
+
200 pcs
3–26