Leveling the Scale Using the Level Indicator
At the place of installation, level the scale using the
leveling feet
(4)
as follows, so that the air bubble is
centered within the circle of the level indicator
(8)
.
Then check to make sure that all four feet are resting
on the surface of the benchtop (stability test).
Extend leveling feet (turn clockwise) to raise the scale.
Retract leveling feet (turn counterclockwise) to lower
the scale.
Note:
When installing the scale on a transportable cart,
a single setting is sufficient.
Equipment used as a legal weighing instrument is not
allowed to be installed on a ramp.
Installing a Drive-on Ramp
(For the models FEP-I, IGP-I, FES-I und IGS-I)
The drive-on ramp must be installed and put into
operation by a trained Sartorius dealer or service
technician.
See the table in the “Accessories” list to order the
drive-on ramp suited for use with your weighing
platform.
1–28