Operating the Balance
Error Codes
Error codes are displayed in the main display or application display for 2 seconds.
The program then returns automatically to the previous mode (e.g., weighing).
Display
Cause
Solution
No segments appear on the display
No AC power is available
Check the AC power supply
The AC adapter is not plugged in
Plug in the AC adapter
Automatic shutoff configured in
Press
e
to turn on the balance
the Setup menu (code
8 7 1
)
or select code
8 7 2
in the Setup
menu (“no automatic shutoff”)
h
The load exceeds the balance’s
Unload the balance
capacity
l
or
err 54
The weighing pan is not in place
Place the weighing pan on
the balance
err 01
Data output not compatible with
Change the configuration in the
> display range
output format
Setup menu
err 02
Calibration/adjustment criterion
Calibrate only when zero
cal.n.possible
not met, e.g.,
is displayed
– not tared
Press
t
to tare
– the balance is loaded
Unload the balance
err 03
Calibration/adjustment could not
Allow the balance to warm up again
cal./adj. interrupt
be completed within a certain time
and repeat the adjustment process
err 06
Built-in calibration weight
Contact your local
int. wt. defective
is defective
Sartorius Service Center
err 07
Function not allowed in balances
Contact your local Sartorius Service
function blocked
verified for use in legal metrology
Center for information on having
the settings changed
err 08
*
The load on the balance is too
Check whether the
<> zero range
heavy to zero the readout
“power-on zero range” is set
err 09
*
Taring is not possible when the gross
Zero the balance
< 0 not allowed
weight is
≤
zero
err 10
The tare key is blocked when there
Press
c
to clear the tare memory;
tare fct. blocked
is data in the tare memory
then you can tare by pressing
t
(e.g., when running the net-total
application); the tare functions
cannot be accessed simultaneously
err 11
Tare memory not allowed
Check the tare value entered
tare 2 blocked
err 12
Tare stored in memory greater than
Check sample/container
tare > max.
weighing range or range limits
err 17
Internal adjustment is not possible
Reduce the preload or change
adj.-wt. >max.
because the preload is too heavy
the configuration
* = occurs only when balance is operated via the SBI interface (ESC f3_/f4_)
Error Codes
Summary of Contents for LP Models
Page 18: ...18 setting not applicable in balances verified for use in legal metrology...
Page 20: ...20 not in balance models with a weighing capacity 16 kg...
Page 81: ...81 Overview Menu Structure setting cannot be changed in balances used for legal metrology...
Page 92: ...92...
Page 93: ...93 Overview...
Page 94: ...94...
Page 95: ...95 Overview Z 14 755 320 34 10 93...
Page 96: ...96...