31
Operating the Balance
Application Programs
Using Verified Balances as Legal
Measuring Instruments in the EU*:
All application programs can
be selected on balances used as
legal measuring instruments.
Calculated values can be indicated
as follows:
Percent
= %
Piece count (counting) = pcs
Computed value
= o,
G
Toggle between Weight Units
Purpose
With this application program
you can switch the display
of a weight value back and forth
between two weight units by
pressing a soft key.
Available Features
– Toggling the displayed weight
– Setting the display accuracy
– Other features as for the basic
weighing function
Factory Settings
Weight unit 1: Grams/g (
1 7 2
)
Display accuracy 1 (in the 1st
range): All digits (
1 8 1
)
Weight unit 2: Grams/g (
3 1 2
)
Display accuracy 2 (in the 2nd
range): All digits (
3 2 1
)
Preparation
The following weight units are available in both ranges:
Unit
Conversion factor
Display
Printout
Grams
1.00000000000
g
g
Kilograms
0.00100000000
kg
kg
Carats**
5.00000000000
ct
ct
Pounds**
0.00220462260
lb
lb
Ounces**
0.03527396200
oz
oz
Troy ounces**
0.03215074700
ozt
ozt
Hong Kong taels**
0.02671725000
tl
tlh
Singapore taels**
0.02645544638
tl
tls
Taiwanese taels**
0.02666666000
tl
tlt
Grains**
15.43235835000
GN
GN
Pennyweights**
0.64301493100
dwt
dwt
Milligrams**
1000.00000000000
mg
mg
Parts per pound**
1.12876677120
o
/lb
Chinese taels**
0.02645547175
tl
tlc
Mommes**
0.26670000000
o
mom
Austrian carats**
5.00000000000
o
K
Tola**
0.08573333810
o
tol
Baht**
0.06578947436
o
bat
Mesghal**
0.21700000000
o
MS
Setting Weight Unit 1
●
Access the Setup menu:
select
Menu
●
Set the parameter for weight
unit 1 (
1 7 2
through
1 7 20
):
See “Configuring the Balance”
●
Set display accuracy 1
(
1 8 1
through
1 8 5
) :
See “Configuring the Balance”
Setting Weight Unit 2
k
Access Setup: select
Menu
●
Set the parameter for weight
unit 2 (
3 1 2
through
3 1 20
):
See “Configuring the Balance”
●
Set display accuracy 2 (
3 2 1
through
3 2 5
):
See “Configuring the Balance“
●
Exit the Setup menu: Press
M
The following levels of display
accuracy are available in both
ranges:
– All digits
– Fewer for weight change
– Last digit off (reduced by 1 digit)
●
Turn on the balance: Press
e
> All segments of the display light
up briefly
●
Configure the “Toggle between
Weight Units” application in the
Setup menu: Press
M
●
Select the configuration mode:
Press
w
(
menu
soft key)
●
Set the parameter
2 1 2
: See
the chapter entitled “Configuring
the Balance”
●
Exit the Setup menu: Press
M
* including the Signatories of
the Agreement on the European
Economic Area
** = not applicable in balances
verified for use in legal
metrology
Summary of Contents for LP Models
Page 18: ...18 setting not applicable in balances verified for use in legal metrology...
Page 20: ...20 not in balance models with a weighing capacity 16 kg...
Page 81: ...81 Overview Menu Structure setting cannot be changed in balances used for legal metrology...
Page 92: ...92...
Page 93: ...93 Overview...
Page 94: ...94...
Page 95: ...95 Overview Z 14 755 320 34 10 93...
Page 96: ...96...