46
Over/Under Checkweighing
O
Purpose
This program is used to check
whether a sample corresponds
to a pre-set target value or is
within a specific tolerance range.
In addition to the display in the
measured value line, the results are
shown on the bar graph and can
also be routed through the interface
port via control lines for further
electronic processing.
Available Features
– Lower tolerance limit (minimum),
target value and upper limit
(maximum) stored in long-term
memory
– Optional balance configuration
in the Setup menu for automati-
cally initializing this application
and loading the values stored in
long term memory for the target
value and the upper and lower
tolerance limits when you turn on
the balance
– Enter target value and limits by
placing a load on the balance or
using the numeric keys
– Control in entering target and
tolerance values, so that the
upper limit
≥
the target
≥
the
lower limit
≥
1 display increment
– Accuracy of a weight readout
or keyboard input as target/
tolerance value stored
corresponds to the display
accuracy
– Optional balance configuration
in the Setup menu for automatic
output to the interface port
(print application parameters)
of target value and tolerance
limits when initialization
is completed
– Control range for the balance’s
data output port lines is 30% to
170% of the target value
– Control range can be set
anywhere from 10% to infinity.
Control lines set when stability is
reached within the control range;
deleted only when you exit this
application
– Optional configuration in
the Setup menu for activation
of control lines independent
of weight value (weight within
control range, stability reached)
– Toggling the display between
weight readout and control
(checkweighing) display
by pressing
v
. If the weight
value exceeds tolerances,
the measured value line shows
the weight while the control
display shows “LL” for “too low”
or “HH” for “too high.”
– Toggle the text line display be-
tween weight value and control
display, nominal value “
setp
” and
tolerance values “
min
” and “
max
”
by pressing the
w
key.
– Weight displayed on bar graph in
relation to upper and lower limits
and target value.
– Optional automatic printout of
weight when it is within the
control range at stability (
auto
print +/-
).
After an automatic printout, the
balance is blocked. Before you
can generate the next printout,
you must unblock the balance
by unloading it (weight must be
under 30% of the target) or
by placing a load on the balance
(bringing the weight up to at
least 170% of the target).
– Press
c
to delete the
initialization parameters and end
the over/under checkweighing
program
Factory Settings
Auto-start application (automatic
initialization with target value and
tolerance limits loaded from
long-term memory): off (
3 10 2
)
Automatic output of target value
and tolerance limits via the interface
port (print application parameters):
off (
7 1 1
)
Auto print +/– (automatic printout
of a weight when it is within the
control range at stability): off (
4 2 2
)
Control lines +/– on
(activating data output port lines):
Within control range (
4 3 1
)
Summary of Contents for LP Models
Page 18: ...18 setting not applicable in balances verified for use in legal metrology...
Page 20: ...20 not in balance models with a weighing capacity 16 kg...
Page 81: ...81 Overview Menu Structure setting cannot be changed in balances used for legal metrology...
Page 92: ...92...
Page 93: ...93 Overview...
Page 94: ...94...
Page 95: ...95 Overview Z 14 755 320 34 10 93...
Page 96: ...96...