47
Operating the Balance
Preparation
The checkweighing program requires
a target value for comparison
to the current value. This target has
a tolerance range, which is defined
by absolute weight values: upper
and lower limits. These limits can
be entered either by storing weights
on the balance or via key input.
There are four control lines,
called data output port lines, which
are activated as follows: (see also
the diagram at the right):
– lighter
– equal
– heavier
– set
The control range spans 30% to
170% of the target value. You can
configure this parameter in the
Setup menu (
4 3
+/- ctrl ports on
)
to select whether the control
lines are:
– always on
– activated within the control range
– activated at stability
– activated at stability within the
control range
This makes it possible,
for example, to connect a simple
indicator for the weighing results
(e.g., three different colors, one each
for the weighing results: too light,
O.K., too heavy).
Response of Control Lines
During Checkweighing
Configurations:
– always on
– activated at stability
Configurations:
– activated within control range
– activated at stability within
control range
Output port specifications
When not in use, the voltage level is high: >2.4 V/+2 mA
When in active use, the voltage level is low: <0.4 V/–2 mA.
!
The output ports are not protected against short circuits!
Summary of Contents for LP Models
Page 18: ...18 setting not applicable in balances verified for use in legal metrology...
Page 20: ...20 not in balance models with a weighing capacity 16 kg...
Page 81: ...81 Overview Menu Structure setting cannot be changed in balances used for legal metrology...
Page 92: ...92...
Page 93: ...93 Overview...
Page 94: ...94...
Page 95: ...95 Overview Z 14 755 320 34 10 93...
Page 96: ...96...