Practical Example
Checkweighing samples of 170 g, with an allowable tolerance of –5 g and +10 g. Printout of upper and lower tolerance limits.
Weighed values printed out automatically when stability is reached and weight is within the control range.
Step
Key (or instruction)
Display/Output
1. Select the checkweighing
see “Preparation”
application in Setup
2. Set the following checkweighing
See “Configuration”
parameters in the Setup menu:
– Auto print +/–: On: values in tol.
– Print app. param.: All parameters
3. Prepare a container for the sample
Place empty container
on the balance
4. Tare the balance
t
5. Enter initialization values
start
soft key
v
6. Store target value (here: 170 g)
Place ideal sample
in container
7. Store target value and
setp
soft key
v
unload balance
Remove ideal sample
from balance
8. Enter value for lower limit
1 6 5
(170 g – 5 g)
plus/minusxxstart
plus/minusxxstart