Net-Total Formulation
R
Purpose
With this application program you can
weigh in different components up to a
defined total. You can also print out the
total weight and the individual weights
of the components.
Available Features
– Taring
– Weighing different components
(maximum: 99 components) from “0”
to a defined total component weight
– Storing component weights
(“Store xx comp.”), with
– display zeroed after value stored, and
– automatic printout (print application
parameters);
either
– of the last component weight
(net value) or
– of the total weight (tare value)
– Display of the transaction counter
“XXth” (referring in each case to the
next component) in the soft key label
display
– Clearing of the component memory
when the weighing series is canceled
c
and printout of the total weight
if you have configured the balance for
GLP-compliant printouts; otherwise,
printout of the net value
p
– Toggling between component weight
and total weight by pressing
w
.
– ISO/GLP-compliant printout
of the total of the individual compo-
nent weights (
Tot.cp
)
Factory Settings
Automatic printout when component
value stored: print net total (
7 3 1
)
Preparation
G
Turn on the balance
> All display segments light up briefly
G
Select the Net Total application
in the Setup menu: Press
M
G
Select menu:
Press the
menu
soft key (
w
)
G
Set parameter
2 1 6
:
See “Configuration”
G
Exit the Setup menu:
Press
M
Setting Parameters for the Net-Total
Application
k
Access the Setup menu: Select
Menu
G
Set the parameter for automatic print-
out when component stored
7 3
Print net total
See “Configuration”
G
Exit the Setup menu: Press
M
Additional Functions
In addition to functions for:
– alphanumeric input,
– taring
(not during alphanumeric input),
– printing,
you can also access the following
functions from this application:
– calibration
(not during alphanumeric input),
– setup,
– turning off the balance.
Calibration and Adjustment
G
Press
q
> See “Calibration and Adjustment”
for further instructions
Setup (setting parameters)
G
Press
M
> See “Configuration” for
further instructions
Turning Off the Balance
G
Press
e
> The balance shuts off
> The display goes blank
40